>>> 'tylan7' problem from bill -- Tue Apr 25 12:42:18 1989 <<<

the overtemp shutoff needs to be calibrated -
the current settings do not allow operation above 1150 C.
I'm trying to perform a 1200 C anneal.
The overtemp setting is currently 850,
shutoff should be setting + 350 = 1200,

however the alarm is sounding at 1130 C and
the circuit breakers are tripping at 1140 - 1150 C.

I can provide some more details if needed - Bill

>>> 'tylan7' problem from hoagland -- Tue Apr 25 16:59:39 1989 <<<

problem:  OT alarm activating below setpoint(dial + 350)- tube
          cannot reach 1200C.

cause:  firing board OT circuit out of adjustment.  Voltage
        reference at 4.6 volts; should be 5.00.

solution:  Voltage reference and OT trim pots adjusted.  Using
           calibration standard OT alarm trips at 1250C; breaker
           trips at 1300C.

>>> 'tylan7' problem from evan -- Thu Aug 17 13:58:43 1989 <<<

ack'ed at 13:55

>>> 'tylan7' problem from evan -- Mon Oct  2 15:32:38 1989 <<<

ack'ed tylan7 at 15:33.  since no one had this enabled, I don't know
who needs to be informed!

>>> 'tylan7' problem from evan -- Thu Oct  5 09:48:09 1989 <<<

An H2/N2 supply line has been routed tto the N2 MFC for a few runs by hing.
It should not be used as normal unttil this temp. modification has been
removed.  tylan7 should be back to normal Friday a.m.

>>> 'tylan7' problem from evan -- Fri Oct  6 08:13:24 1989 <<<

The H2/N2 modification has been removed and the normal N2 MFC is back online.
tylan7 is up for standard operations.

>>> 'tylan7' problem from hoagland -- Tue Nov  7 10:09:46 1989 <<<

There seems to be a large piece of gasket material sticking
out of the stainless steel flange assemble in front of tube7.
If the integrity of the tube seal has been violated(not likely)
this might lead to attack on metal parts around the tube during a
TCA clean.