>>> 'tempress' problem from carlos -- Wed Jan 31 22:51:17 1990 <<<

Level problem and indexing problem => caused the blade to break

Once the height was set to 14 mils below the reference, I proceded
to do the azimuth alignment. The blade did not touch the wafer 
while I did the alignment (as it is supposed to do) near the 
center of the wafer, but once everything was set, I moved the 
blade all the way in to the first cut and several things happened:

	1. The blade touched the chuck and was damaged
	2. The indexing would not go in all the way, and 
	   I could not make the first cut.

Maybe there is a chuck leveling error causing 1. and something wrong
with the index causing 2.

The blade needs to be replaced.

>>> 'tempress' problem from cleland -- Fri Feb  2 19:59:43 1990 <<<

The blade cut is right on the edge of the view screen, out of range of
adjustment with the y-axis fine control. It needs to be moved up to
line up with the crosshair. Otherwise, no problems.

>>> 'tempress' problem from carlos -- Sun Feb  4 15:42:58 1990 <<<

It works fine except that the cut is made about half a screen below 
the cut line.

>>> 'tempress' problem from phillip -- Mon Feb  5 08:10:36 1990 <<<

P: Optical hair line askew relative to blade cut on wafer.
S: Replaced view window scope bulb. Aligned scope to hairline
   cut. The tempress is up.

>>> 'tempress' problem from costello -- Wed Feb  7 16:34:31 1990 <<<

saw cut is above the field of view of the microscope.

>>> 'tempress' problem from phillip -- Thu Feb  8 11:28:18 1990 <<<

P: Microscope hair line 'off' from cut .
S: Refastened loose allen screw (left loose
   from last adjustment....oops!)
The saw is fixed and available for use.

>>> 'tempress' problem from nam -- Fri Feb  9 11:07:59 1990 <<<

blade is broken.

>>> 'tempress' problem from phillip -- Fri Feb  9 11:53:18 1990 <<<

Replaced tempress blade per Nam Kim's report .

>>> 'tempress' problem from liew -- Tue Feb 20 10:10:56 1990 <<<

The saw leaves a milky brown stuff along the cut. 

>>> 'tempress' problem from bob -- Tue Feb 20 11:47:56 1990 <<<

The milky (brown) patches left on liew's wafer were the result of 
insufficient cooling water. The valves that normally control this
flow were in there proper setting and, in fact had been cemented
there with RTV. The main shutoff valve was found only partially
open. It is not used  to regulate flow and should have been fully
open.

The water flow has been reset and the saw is now cutting without
leaving a residue. Phill will recement these valve settings upon
his return. In addition the 8 amp control fuse needs replacing.
It is a 15 amp fuse and the machine requires an 8 amp fuse.
This does not effect users.

The tempress saw is up.

>>> 'tempress' problem from liew -- Tue Feb 20 12:34:56 1990 <<<

Water not draining fast enough and overflowed to table. Solution: I stopped 
for a moment after each cut.

>>> 'tempress' problem from bob -- Wed Feb 21 16:08:33 1990 <<<

ThThe water has been decreased on the tempress. This shold prevent
the tray from overfilling.
The tempress is up.

>>> 'tempress' problem from kim -- Fri Feb 23 13:36:30 1990 <<<

Problem:  Water leaks from underneath the Tempress Saw.

>>> 'tempress' problem from phillip -- Tue Feb 27 06:45:44 1990 <<<

Replaced small orifice drain grates with larger diameter orifice
drain grates. The problem as reported by Kim about excess water
is fixed.

>>> 'tempress' problem from jaytu -- Tue Feb 27 22:58:50 1990 <<<

I experienced the indexing problem when running the saw.  Also, the water
leak reappeared.  No water drained into the bottle at all.

>>> 'tempress' problem from jgtsay -- Wed Feb 28 01:51:10 1990 <<<

what happen to the water?

>>> 'tempress' problem from scotts -- Wed Feb 28 17:58:32 1990 <<<

the index on channel 1 at least does not seem to be self-consistent.  I made several cuts in a sample then scanned (using the index) back across my wafer and
the cuts did not!! line up with the video alignment mark. 

>>> 'tempress' problem from murray -- Sun Mar 18 17:35:36 1990 <<<

	When the saw initially was turned on on I heard a vibration, which disdisappeared after a few moments.  Believing that everything was all
right, I proceeded to scribe my wafer.  After the saw had completed about 
half of the wafer the saw jammed.  Taking the cover off of the blade,
I noticed that the saw was unable to rotate and that the blade was broken.
I turned the instrument off and am reporting the problem.

>>> 'tempress' problem from murray -- Sun Mar 18 17:56:14 1990 <<<

	The tempress siezed half-way thru cutting a wafer. Upon inspection
I found that the saw was stuck (could not turn w/ my hand) and that
the blade appeared to be broken. I had a user named "Carlos" inspect
the saw, and unable to find a simple explanation for the problem, turned
the instrument off and reported the problem.

>>> 'tempress' problem from phillip -- Mon Mar 19 09:06:25 1990 <<<

P: Spindle stalled?
S: Replaced blade. Dressed blade.
   The cause of the problem over the weekend is unknown.

The tempress saw is up and available for use.

>>> 'tempress' problem from jaytu -- Mon Mar 19 22:22:32 1990 <<<

Water did not drain even after 1 cut.  Leaked after every cut.
Also, after rotating by 90 degrees for the cross cuts, the cooling
water started turning brown under the saw.

Alignment problems were not severe until after the 90 degree rotation.
After the rotation, while the saw was in dice mode, the saw made 2 misaligned
cuts in one pass through.

>>> 'tempress' problem from guochun -- Thu Mar 22 23:10:49 1990 <<<

TThe rotator of the blade is jamed! It doesn't rotate AT ALL.
guochun

>>> 'tempress' problem from bob -- Fri Mar 23 12:18:33 1990 <<<

There is a problem with the air bearing spindle of the tempress.
Repairs will take several days. Users will be informed of the progress.

>>> 'tempress' problem from murray -- Sun Mar 25 18:21:04 1990 <<<

Blade on tempress not spinning.

>>> 'tempress' problem from phillip -- Mon Mar 26 06:35:33 1990 <<<

P: The spindle will not spin.
S: The inevitable has happened. The air bearing spindle
   has locked up and cannot be freed. The only way to
   remedy this problem is to replace this $7000.00+ part.


        THE TEMPRESS SAW IS DOWN INDEFINITELY!

>>> 'tempress' problem from phillip -- Tue Mar 27 14:05:41 1990 <<<

P: Spindle not turning.
S: Mechanician Jose Rivera repaired (and aligned)
   air bearing portion of saw. The saw is up and 
   available for use.

>>> 'tempress' problem from hebert -- Wed Mar 28 12:05:10 1990 <<<

The spindle locked up once while cutting a 100mm wafer. Phil loosened
the drive belt. After this, the spindle turned freely. The spindle
drive motor gets very hot.

The tempress is UP.

>>> 'tempress' problem from murray -- Thu Mar 29 14:45:08 1990 <<<

	The blinking light does not come on when the platform is
raised to adjust depth of cut.  Instrument is inoperable in this 
condition.
>>> 'tempress' problem from murray -- Thu Mar 29 14:47:25 1990 <<< [No Comment]

>>> 'tempress' problem from bmartin -- Thu Mar 29 19:08:46 1990 <<<

The diamond saw does not go to the correct cutting depth, and the
y-scan only moves in one direction after the height is adjusted.
>>> 'tempress' problem from xiaohong -- Fri Mar 30 10:23:31 1990 <<< [No Comment]

>>> 'tempress' problem from jgtsay -- Fri Mar 30 13:57:47 1990 <<<

The height adjustment doesn't work. Ref light won't flash.

>>> 'tempress' problem from voros -- Sat Mar 31 11:19:19 1990 <<<

The tempress is down waiting for repair. Pls do not use.

>>> 'tempress' problem from phillip -- Mon Apr  2 10:34:58 1990 <<<

P: Sensor for 'chuck height' not functioning.
S: Replaced resinoid blade with lab standard
   diamond blade. Burnished height sensor contact
   on tail end of air bearing mechanism.
   Installed and dressed new diamond blade.
The tempress is up and available for use.

>>> 'tempress' problem from dchin -- Mon Apr  2 16:57:25 1990 <<<

The optical microscope on the tempress saw could not be focused on wafers.
The features on the wafers could not be seen. When I turned the nob on the
scope to adjust the focus, it slipped.

The saw itself works. Indexing seems OK too.

>>> 'tempress' problem from phillip -- Tue Apr  3 10:33:32 1990 <<<

P: Microscope not easily focused.
S: Repositioned scope illumination bulb. 
   The saw is up and available for use.

>>> 'tempress' problem from bob -- Mon Apr  9 15:19:39 1990 <<<

User reports low water flow/wafers covered with debris.
The water cooling on the tempress has been increased
to sufficient levels. It is likely the control valves for
this are vibrating closed. These will be glued at the
appropriate settings by Phil Guillory tomorow.
The tempress is up for use in the meantime.

>>> 'tempress' fix from phillip -- Wed Apr 11 08:02:31 1990 <<<

P: Cooling water backs up drain and tends to spill
   onto floor.
S: Ran separate internal drain from cooling water
   drain tray to external floor drain. This fixes
   a long standing problem.
The tempress is up and available for use.

>>> 'tempress' problem from liew -- Sat Apr 14 16:16:43 1990 <<<

The saw cut too deep and actually cutting the bottom plate for
a depth setting of 0.0065"

>>> 'tempress' problem from murray -- Sun Apr 15 14:23:34 1990 <<<


	Y-axis control will not work for me.

Matt

>>> 'tempress' fix from phillip -- Mon Apr 16 07:47:31 1990 <<<

P: Y--axis control doesn't work per Murray
S: Replaced damaged blade.
   The y--axis fine adjust knob cannot be turned
   during a cut. This action ruins the blade.
   The saw is up and available for use.

>>> 'tempress' problem from dchin -- Wed Apr 25 01:44:37 1990 <<<

The microscope focus does not work. The knob for adjusting focus slips when
it is turned.

>>> 'tempress' fix from phillip -- Wed Apr 25 08:03:41 1990 <<<

P: Scope optics out of sinc with actual blade cut per dchin.
S: Successfully diced 2" wafer wiith the specs Dave had left
   dialed into saw controls. NO problem with optics was encountered
   during or after (at least) 10 cuts.
No trouble found. The saw is up and available for use.

>>> 'tempress' problem from xiaohong -- Fri May  4 12:32:44 1990 <<<

esWhen press HEIGHT, move chuck up, the REF light doesn't flashes or lits.
The blade on the machine is resinoid blade.

>>> 'tempress' fix from phillip -- Fri May  4 13:21:38 1990 <<<

Will replace broken resinoid (or diamond) cutting wheel
after need verification. The problem with the tempress saw
will be remidied shortly.
.,

>>> 'tempress' problem from xiaohong -- Wed May  9 15:27:42 1990 <<<

When I used resin blade to cut sapphire wafer. The blade alwayse broke.
Today two blades broke. For the 1st blade, I set HEIGHT 0.004 inch and
SPEED low 6, after cut 1/2 line it broke. For the 2nd blade, I tried to
cut sillicon wafer first, it worked well. Then I cut sapphire, this time 
I set HEIGHT 0.003 inch (my wafer is 17 mil thick) , it only cut 2 lines.
Does anybody can help me to solve the problem? Or does anyone has the 
experience to cut sapphire ? Please let me know. Thanks!

Xiaohong

>>> 'tempress' fix from bob -- Mon May 14 11:38:43 1990 <<<

New resin blades have been ordered for the tempress saw. This will solve
a breakage problem when cutting saphire. The tempress is up and available
for use with it's regular blades.