>>> 'nanoline' problem from bob -- Wed Oct  4 14:07:27 1989 <<<

The light source on the nanoline has been broken and needs replacing.
The broken part is a plastic composition tube and will be ordered from
Nanometrics.

>>> 'nanoline' problem from hoagland -- Thu Oct  5 16:16:02 1989 <<<

An effort to reconnect the lamp coupler which separated earlier
this week failed- the locktight super glue did not harden.
A new coupler will be ordered tomorrow.

>>> 'nanoline' problem from hoagland -- Tue Oct 10 08:33:30 1989 <<<

A workorder has been submitted to the machine shop to have the
broken coupler for the nanoline light source replicated and
replaced by an aluminum part.  As soon as that work is complete
the nanoline will be back up and available for use.

>>> 'nanoline' problem from hoagland -- Mon Oct 16 10:55:06 1989 <<<

An aluminum coupler has been installed on the nanoline light source
to replace the broken plastic part.  The nanoline is up and available
for use.

>>> 'nanoline' problem from kim -- Tue Apr 24 15:11:46 1990 <<<

Missing 100X objective.

>>> 'nanoline' problem from kim -- Fri Apr 27 15:13:24 1990 <<<

Problem:  Autofocus doesn't work sometime---because of vibration?

>>> 'nanoline' fix from kim -- Thu May  3 07:48:05 1990 <<<

Problem:  Autofocus fails sometime.
Cause:    Faint images or rought edges.
Solution: Change to the faint or rough measuring program.