>>> 'iv' problem from evan -- Thu Jan 26 15:57:03 1989 <<<

problem: noise in cables
cause: old, bent banana connectors; lack of grounding
solution: replaced all banana plugs using color coded sheaths(red,green,yellow)
added a ground strap from the scope ground to a banana jack on the back of the
column. This grounds the base and microscope as well as the shields of the
internal interconnecting wires (these had been ungrounded and could have been
a source of some of the noise).
Added another ground from the back of the column to the chrome plate (this is
insulated from the base). This can be unpluged if desired.
The test stage is still isolated. The existing lead (to the orange plug) is 
still intact.

>>> 'iv' problem from khui -- Mon Jan 30 13:37:32 1989 <<<

one of the probe tip is slightly bended when I enabled the machine.  No
big problem.

Kelvin.

>>> 'iv' problem from carlos -- Thu Feb  9 00:27:24 1989 <<<

there is no vacuum

>>> 'iv' problem from hoagland -- Thu Feb  9 16:45:36 1989 <<<

problem:  User reports no vacuum for IV.

cause:  Unknown.  IV station was checked and found to have vacuum.

solution.  None.  Users should be aware there are 2 vacuum switches-
           one to secure the probe assemblies, and one to secure the
           sample wafer.

>>> 'iv' problem from voros -- Mon Mar  6 10:54:32 1989 <<<

Taurus box does not turn on when iv is enabled on computer

>>> 'iv' problem from carlos -- Fri Mar 31 03:59:38 1989 <<<

The left micromanipulator is jammed

>>> 'iv' problem from bob -- Fri Mar 31 07:46:41 1989 <<<

The micromanipulator was unjammed and the set screw on the knob
retightened. Manipulators should not be run to their extreme
limits and then have torque applied.
The iv is up.

>>> 'iv' problem from kuchta -- Fri Jun 16 14:54:01 1989 <<<

The voltage knob on the curve tracer has come loose.

>>> 'iv' problem from bob -- Fri Jun 23 10:14:12 1989 <<<

The knob on the iv probe station has been replaced as it cracked.

>>> 'iv' problem from voros -- Fri Jul  7 10:54:22 1989 <<<

Stage is lose, y movement is slipping.

>>> 'iv' problem from hoagland -- Mon Jul 10 11:46:34 1989 <<<

problem:  Y axis stage motion is loose; stage exhibits almost
          a 1/4 in. of play.

cause:  Loose set screw on Y axis worm gear stage termination.

solution:  Tighten set screw.

>>> 'iv' problem from kuchta -- Thu Jul 13 19:49:47 1989 <<<

Minor problem with curve tracer:  Plastic peak power disk has broken off
of its knob so one does not really not what the current peak power setting
is.

>>> 'iv' problem from hoagland -- Tue Jul 18 11:54:40 1989 <<<

The curve tracer is being repaired today and should be up
again this afternoon if all goes well.

>>> 'iv' problem from hoagland -- Tue Jul 18 18:22:40 1989 <<<

Repair of the curve tracer will take slightly longer than 
expected because of mechanical difficulties realigning the 
collector high voltage switch.  With luck it should be up
sometime tomorrow.

>>> 'iv' problem from hoagland -- Thu Jul 20 08:55:44 1989 <<<

The failure to repair this equipment stems from a knob on the
maximum collector switch which failed to detach as intended by
the manufacturer.  This knob was to have a broken indicator
skirt reattached with glue.  To remove the knob and shaft assembly
(now required) necessitated pulling apart the internal shaft coupler
for both inner and outer shafts of the switch.  The knob was
repaired successfully, but lining up and securing the shaft
couplers later proved all but impossible(as suggested by
Tektronix).  Therefore a replacement curve tracer will be
installed shortly in place of the now useless unit.

>>> 'iv' problem from bob -- Fri Jul 21 13:45:24 1989 <<<

A new curve tracer has been installed next to the iv. Steve will
check the cables to make sure it interfaces easily.

>>> 'iv' problem from hebert -- Mon Jul 31 12:30:41 1989 <<<

The probe tips are bent.

>>> 'iv' problem from robin -- Mon Jul 31 12:57:42 1989 <<<

P: Bent Probes
C: Usage
S: Replace probes.

>>> 'iv' problem from carlos -- Fri Aug  4 00:36:23 1989 <<<

the vacuum of the probe manipulators is not working

>>> 'iv' problem from gian -- Fri Aug  4 09:37:21 1989 <<<

Vacuum problem is fixed...the rear probe base was dirty and leaking.
-David Giandomenico

>>> 'iv' problem from denlinge -- Sat Aug  5 18:57:25 1989 <<<

the probe needles are badly adjusted, the right (green) probe runs
out of range in the up direction at almost the same place as the
left (blue) probe runs out of room in the down direction, this makes
it difficult to raise and lower them on the sample.  I am new to the
machine so I did not want to adjust the needles myself.

>>> 'iv' problem from denlinge -- Wed Aug  9 13:55:37 1989 <<<

Theta control on the stage does not work in the counter-clockwise
direction.
>>> 'iv' problem from denlinge -- Wed Aug  9 16:33:08 1989 <<< [No Comment]

>>> 'iv' problem from hsin -- Fri Sep 22 14:55:59 1989 <<<

I found the right hand probe can not raise up or lower down.
Some one must broke it without report.  Also the two probes
remained on machine both have a curved point,i.e. the tips
are also damaged.

Wei Hsin

>>> 'iv' problem from mats -- Sun Sep 24 18:14:20 1989 <<<

1:  The problem reported by Hsin is nonexistent: all three probes work.
    However, somebody had turned one manipulator about 40 turns (!!) out
    of range, until the screw disengaged...

2:  The sample XY-stage has developed an annoying amount of play. This
    appears to be because the x-motion rails need to be tightened.

>>> 'iv' problem from evan -- Mon Sep 25 10:47:58 1989 <<<

problem: xy-stage has excessive play
solution: checked adjustment of x and y slider rails and found them to be
normal.  Found the screws that hold the top of the stage (and y-direction rails)to the x-direction mechanics (on underside of main platform) to be loose.  These
were tightened and the excessive play was removed.  These screws are accessed
through the hole in the y-platform behind the work platform riser shaft. There
are two screws and the y-stage must be moved in order to align the hole with
screw heads.

the iv is functional

>>> 'iv' problem from kuchta -- Wed Nov  8 16:07:16 1989 <<<

stage rotation knob has come loose.  The iv is function otherwise.

>>> 'iv' problem from carlos -- Tue Nov 14 12:16:24 1989 <<<

out of new probe tips

>>> 'iv' problem from gian -- Fri Dec  8 12:25:59 1989 <<<


The microscope light flickers intermittently, especially at the lowest
setting.

>>> 'iv' problem from hoagland -- Wed Dec 13 11:50:12 1989 <<<

problem:  User reports microscope illuminator flickers.

cause:  Bad contacts in bulb socket.  Heat caused oxidization
        of socket connector and bulb pins.

solution:  Socket contacts and bulb pins burnished.  Illuminator works
           fine now- the iv station is up.

>>> 'iv' problem from mats -- Sat Mar  3 01:06:19 1990 <<<

A minor one:  The Y translation stage for the wafer chuck needs to be
tightened (I can see, but not reach, the four set screws in question).
It works fine as it is, except that the excessive play lets the wafer
chuck vibrate, which makes micron scale work tricky.

>>> 'iv' problem from tom -- Thu Mar  8 15:15:18 1990 <<<

Problem:      User reports vibration and play on the Y control.
Cause:        Loose knob on Y  control.
Solution:     Tightened knob. There is still some vibration but the 
              back left air puck is not inflated. I do not have time to
              find the atatchment to fill it. Later, later.....