>>> 'gases' problem from phillip -- Wed Jan 17 14:11:04 1990 <<<

Loaned labuser James Yeh one of three gas carts. He will return it
when he is done with it (this week).

>>> 'gases' problem from evan -- Tue Jan 23 08:50:13 1990 <<<

The size 2 cylinder of neon is in the lab. Cyl # FF25589.
James, see me for arranging hookup.

>>> 'gases' problem from bob -- Mon Feb 19 11:28:33 1990 <<<

Alarm on the O2 were caused by a LOX cylinder that wasn't boiling
O2 fast enough, (to good a dewar!). I valved on the reserve
cylinder whitch restored the pressure to 40 psi. I can opn
the pressure building valve on the errant LOX bottle;
however, this will increase the boil off rate to a point
where we are waisting O2. We may do this upon return but for
now we are on a full bottle at the proper pressure and the
low pressure bottle will act as a backup. Worst case, we
deliver O2 at 20 psi which is still capable of supporting
the Microlabs needs.

Evan, lets look at this situation together, we may have a
leaky chack valve on this cylinder or we may be just on the
edge of having enough boil off due to this cylinder being a
very good dewar!

>>> 'gases' problem from bob -- Wed Feb 21 07:10:34 1990 <<<

The Ar cylinder on heatpulse1 needs changing.

>>> 'gases' fix from phillip -- Tue May  1 13:33:14 1990 <<<

Replaced H2 gas bottle on H2/O2 torch next to the ionmill.
There was no relative empty tank so a full bottle was placed on line.
(There is no hydrogen currently in the gas storage room. Working
to have some delivered by receiving (2nd request).)